Testing Software to Prove It Works: How Fully Validated Devices Ended up in 2025 Class I Recalls

Chris explains why fully validated devices still face Class I recalls and how cross-functional scenario testing exposes missed software and use-environment risks before V&V.

Plays from YouTube when you press play. Cookie settings

What the session covered

In this session, host Sean Smith was joined by Chris of Keystone Medical Device Partners to examine why fully validated medical devices still appear in Class I recalls. Chris began with FDA data presented by Ann Reed showing a 232% rise in Class I recalls from 2020 to 2025, including 111 in the previous year, and noted that software was a recurring theme in audience suggestions. He clarified the difference between a Class I recall, the most severe category, and a Class I device, which is not high risk. Chris then walked through recent public recalls, including a nitric oxide delivery system that stopped mid therapy when a user pressed two on-screen buttons in quick succession, and an infusion pump that stopped when hospital network settings changed. Neither scenario had been captured in requirements or testing. The main part of the session described a KPMG cross-functional scenario workflow. Teams map critical paths, gather recall and defect history, involve clinicians and hospital IT, generate failure scenarios across disciplines, prioritise them by impact and probability, and convert them into test cases. Chris stressed that the process does not replace hazard analysis, FMEA, human factors validation or cybersecurity threat modelling. It uses their outputs as inputs. The takeaway for viewers is that validation against requirements is not enough if reasonable use and environment are missing from those requirements.

Figures

Figures Kris cited

10 yearstime Chris spent running software testing for medical device manufacturers
18 monthslength of the enterprise systems program with Class 3 infusion pumps
6European hospital sites where the infusion pump program went live
232%increase in Class 1 recalls from 2020 to 2025
111Class 1 recalls in the last year
70%share of North American credit card transactions routed through Innovative Merchants' system
40people brought into the cross-functional scenario session
5,000pumps in the hospital dashboard example
500docking stations supporting those pumps
20scenarios each participant was asked to contribute
7questions in the self-assessment
50places limited in the sprint cohort

Source: Sean

Chapters

Jump to a moment

  1. Welcome, upcoming events, and Chris Kelly's background
  2. Class 1 recalls are rising and audience poll on causes
  3. Cybersecurity testing versus software V&V expectations
  4. Recall examples: nitric oxide delivery and infusion pump network settings
  5. Silos, the gap between teams, and why an outside facilitator runs the process
  6. Running the 40-person cross-functional scenario session
  7. Prioritizing scenarios and turning them into test cases
  8. Performance testing, what the method does not replace, and audience Q&A

Times run from the start of the recording.

Take these with you

Six checks

Map critical paths firstStart from the PRD and identify the workflows and interfaces that cannot fail before generating failure scenarios.
Include every disciplineBring engineering, clinical, usability, tech support, tech writing, marketing and program management into the same scenario session.
Do homework before the sessionReview the recall database, the company's defect management system, clinician input and hospital IT practices before generating scenarios.
Prioritize by impact and probabilityPlace scenarios into high or low impact and probability quadrants and focus testing on the high-impact items.
Test the system around the deviceInclude hospital network changes, connectivity loss, battery limits and docking-station behavior in test plans.
Keep existing disciplines intactDo not replace hazard analysis, FMEA, human factors validation or cybersecurity threat modeling; use their outputs as inputs.

Full recording

50:52

Plays from YouTube when you press play. Cookie settings

How to cite this session

Kelly, K. (2026, September 25). Testing Software to Prove It Works: How Fully Validated Devices Ended up in 2025 Class I Recalls [Webinar]. MedTech Leading Voice, Friday In-Focus. medtechleadingvoice.com/watch/2026-09-25-testing-software-to-prove-it-works-how-fully-validated-devices-ended-up-in-2025-class-i-recalls

Earlier sessions

13 Nov 2026Architecting a Claims Support System 6 Nov 2026Methodological Quality and Scientific Validity Under the MDR 30 Oct 2026AI Pre-Submission Review Under the MDR

Latest issues

Issue 191Device Cybersecurity Is Now Post-Market Surveillance Issue 190FDA and the Future of Generative AI Medical Devices Issue 189MDR and IVDR Evidence Requirements in Transition