What the session covered

The webinar opened with host announcements about upcoming MedTech Leading Voice events, including an FDA AI/ML submission talk, UDAMED deadlines, the International MedTech Safety Conference, and a BSI roadshow. Sean then introduced Marcus, head of post-market surveys at Cody Pasco, and handed over for the main presentation. Marcus explained how PMCF surveys fit under the MDR. He compared real-world data with traditional clinical studies, described PMCF as a continuous process that updates the clinical evaluation, and reviewed data collection methods such as clinical studies, registries, surveys, complaints, and literature. He contrasted high-quality level 4 surveys with more general level 8 surveys, covering when each is appropriate and why surveys are not suitable for every device. He then discussed common problems: poor planning, weak endpoints, low response rates, recruitment difficulty, bias, and limited longitudinal data. He outlined best practices including a formal survey protocol, statistical sample size justification, validated survey software, respondent validation, local-language translation, quality controls, and transparent reporting. Audience questions covered notified body relationships, language handling, and use of surveys for Class 3 devices.

Figures

Figures Marcus cited

600post-market surveys delivered by Marcus and his team
15–20 minutesmaximum target length for surveys
seven and a half yearstime Marcus has led the post-market survey team
2015year Marcus joined Cody Pasco
2018year the team began focusing on PMCF surveys

Source: Marcus

Chapters

Jump to a moment

  1. Welcome and upcoming MedTech Leading Voice events
  2. Introduction to Marcus and Cody Pasco
  3. Real-world data versus traditional clinical studies
  4. Where PMCF fits in the MDR
  5. PMCF data collection methods
  6. Level 4 versus level 8 surveys
  7. Survey challenges and audience Q&A
  8. Designing high-quality PMCF surveys

Times run from the start of the recording.

Take these with you

Six checks

Justify the survey firstDecide whether a survey is appropriate for the device and document the rationale before choosing it as a PMCF method.
Write a survey protocolSet out device information, objectives, endpoints, inclusion criteria, sample size, and recruitment plan in a formal protocol.
Anchor questions to endpointsMake sure every survey question ties back to the safety and performance endpoints you set in the protocol.
Plan recruitment and languageMap your user base, choose target markets, and plan local-language translation and respondent validation before launch.
Use validated tools and QCUse validated survey software, user acceptance testing, and quality checks so the data and audit trail are defensible.
Report honestlyInclude negative findings, limitations, and protocol deviations in the survey report rather than presenting only favourable results.

Full recording

52:38

Plays from YouTube when you press play. Cookie settings

How to cite this session

Torr, M. (2026, April 10). PMCF Surveys Under MDRWhen They Apply, How to Structure Them, and WhatHigh-Quality Looks Like [Webinar]. MedTech Leading Voice, Friday In-Focus. medtechleadingvoice.com/watch/2026-04-10-pmcf-surveys-under-mdrwhen-they-apply-how-to-structure-them-and-whathigh-quality-looks-like